Type 3595 is a two-microphone probe kit for measuring sound intensity, specifically designed for use with 2260 Investigator™ Hand-held Sound Intensity System.
The probe set includes the 1/2" Sound Intensity Microphone Pair Type 4197 enabling 1/3-octave centre frequency measurements between 20 Hz and 6.3 kHz.
Extending the upper 1/3-octave centre frequency to 10 kHz can be achieved using pressure correction.
- Uses
- Sound intensity measurements using two-microphone technique, in accordance with IEC 1043 Class 1
- Sound power measurements in accordance with ISO 9614-2, ECMA 160 and ANSI S 12-12
- Features
- Microphone pair matched for phase and amplitude response
- Individual calibration data
- 1/3-octave centre frequency ranges
- 20 Hz to 10 kHz with corrections
- 50 Hz to 6.3 kHz in accordance to IEC 1043 class 1
- Minimal shadow and diffraction effects
- Well-defined acoustical microphone separation
- Specially designed for use with 2260E Hand-held Sound Intensity System
|
Used with 1/2" Microphone pair Type 4197, the probe complies with IEC 1043 Class 1. These 1/2" microphones feature patented phase-corrector units making precision low-frequency phase matching a practical possibility, leading to increased measurement range and accuracy
To measure sound intensity accurately using a two-microphone technique, you need a reliable sound intensity probe set containing a matched microphone pair to obtain information on both the instantaneous pressure and pressure gradient in the sound field. The microphones are separated by a fixed distance in the sound field, and the microphone signals are fed to a sound intensity processor which calculates the sound intensity. The sound intensity is calculated from the time average of the sound pressure multiplied by the particle velocity (calculated from the measured pressure gradient). Such a system measures the component of the sound intensity along the probe axis and also indicates the direction of energy flow
Probe Description
The sound intensity probe is constructed on a face-to-face design. It consists of a robust frame which holds the microphone preamplifier(s) and matched microphones in a face-to-face configuration. The distance between the microphones is defined by solid plastic spacers which are held in place by threaded studs on the microphone grids. Sound is constrained to act on each microphone through a narrow slit between the spacer and the microphone grid. This arrangement gives well-defined acoustic separation of the microphones and minimises shadow and reflection effects.
The probe is strong but lightweight and can be held using either a handle with integral cable ≈ 2 m long, ending in a 10-pin Lemo plug or with an extension rod. The probe can be connected to 2260 Investigator via a cable or an extension rod. All the probe sets are supplied in transport cases containing a microphone pair, windscreen (ellipsoidal) and accessories.
The case has pockets to accommodate 2260 Investigator, a Sound Level Calibrator Type 4231 and other small accessories.
Type 3595 is supplied with 1/2" Sound Intensity Microphone Pair Type 4197. These microphones operate on a polarization voltage of 200 V
Sound Intensity Microphone Pairs
Phase matching of the 1/2" microphone pair Type 4197 is better than 0.05 ° between 20 Hz and 250 Hz, and is better than f/5000 degrees at higher frequencies, where f is the frequency. Such phase matching is possible as a result of the integral microphone phase-corrector units (patented) which are fitted to the 4197 microphones. The normalised microphone frequency responses differ by less than 0.2dB up to 1 kHz and by less than 0.4 dB up to 7.1 kHz
Type 4197 is supplied with 8.5 mm, 12 mm and 50 mm spacers. Calibration data provided include phase matching up to a 1/3-octave centre frequency of 6.3 kHz, microphone sensitivities at 250 Hz, actuator responses and individual free-field frequency responses valid for the microphones mounted on a 1/4" preamplifier.
manufacturers' specifications
|
| B&K 4197 | Microphone Pair including UC 5349: 8.5 mm spacer, UC 5269: 12 mm spacer and UC 5270: 50 mm spacer |
| B&K 2683 | Dual Preamplifier |
| B&K AO 0522 | Adaptor Lemo for Jack plug |
| B&K DP 0888 | Coupler |
| B&K HT 0015 | Earphones |
| B&K UA 0781 | Ellipsoidal Windscreen |
| B&K UA 1439 | Extension Stem |
| B&K UA 1440 | Handle with Integral Cable |
| B&K KE 0379 | Carrying Case |
| B&K 4231 | Acoustic calibrator - class 1 |
| B&K 4297 | Sound Intensity Calibrator |
| B&K AO 0441 | Cable - probe extension - 10-pin Lemo |