Manufacturers' Specifications
Charge Amplifier Input : 10-32 UNF socket and BNC socket Sensitivity Conditioning : 0.1 to 10.99 pC/ms-2 3 x digit dial up switches. Amplifier Sensitivity : 0.01 mV to 10 V/pC corresponding to -40 to +80 dB with transducer capacitance of 1 nF Calibrated Output Ratings : Selectable in 10 dB steps Acceleration : 0.1 mV to 1 V/ms-2 Velocity : 10 mV to 100 V/ms-1 Displacement : 0.1 mV to 10 V/mm Signal Output : 10-32 UNF socket and BNC socket Max Output : 8 V peak Output Impedance : < 1 Ω DC Offset : < ± 50 mV Frequency Ranges : -10 % limits quoted Acceleration : 0.2 Hz - 100 kHz or 2 Hz - 100 kHz Velocity : 1 Hz - 10 kHz or 10 Hz - 10 kHz Displacement : 1 Hz - 1 kHz or 10 Hz - 1 kHz Filter - Low Pass : Switchable - 10 % frequency limits of 100 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz and > 100 kHz with attenuation slope of 40 dB/decade Inherent Noise : 2 Hz to 22 kHz 0.005 pC referred to the input with maximum sensitivity and 1 nF transducer capacitance Test Oscillator : 160 Hz sinusoidal, factory preset for test level of 1 V Overload Indicator : Overload LED light when input or output of amplifier is overloaded by high peak levels Rise Time : ~ 2.5 V/µs Environmental : Temperature Range : -10 to +55 °C Humidity : 0 to 90 %RH, no condensation Internal Power Supply : Batteries : 3 x 1.5 V, IEC LR20, alkaline cells Battery Life : 100 hours, approximately Dimensions : height = 133 mm, width = 70 mm and depth = 200 mm Weight : 1.45 kg including batteries
B&K Help Line - 01438 739 000
|