Gracey Hire Logo
Gracey & Associates Hire and Calibration

Bruel & Kjaer 4515-B-0002 Specifications


Bruel and Kjaer 4515


 B&K 4515 Triaxial Seat Pad Accelerometer £ 10 a day * 

Deltatron Triaxial Seat Pad including a Bruel & Kjaer 4524-B accelerometer

Dynamic Characteristics
Voltage sensitivity : 10 ± 5% mV/ms-2 : 100 mV/g +3 -7%   **
Measuring Range : ± 500 ms-2 : ± 50 g
Frequency Response : 0.4 Hz to 900 Hz - typical.
Mounted Resonance Frequency : > 2700 Hz
Amplitude Response : 0.4 Hz to 900 Hz ± 10%
Transverse Sensitivity : < 5%

Electrical Characteristics
Output Impedance : < 30 ohm
DC Output Bias Voltage : + 13 ± 1 V DC

Power Requirements : all three axis must be powered during operation.
Power Supply, unloaded : 24 to 30 V DC
Constant Current : 2 to 10 mA
Warm-up time, 90% of stabilized bias : 10 s

Environmental Characteristics
Temperature Range : -10 to +70 degree C
Humidity : Hermetic
Max. Operational Sinusoidal Vibration (peak) : 5000 m/s² : 500 g
Max. Operational Shock (peak) : 50000 m/s² : 5000 g
Thermal Transient Sensitivity : 0.1 m/s²/ °C : g/°F
Magnetic Sensitivity : 50 Hz - 0.03 Tesla : 20 m/s²/T : 2 g/T

Physical Characteristics
Dimensions : see outline drawing above
Weight : 345 gram
Seat Pad Material : Oil-resistant Nitrile Rubber, Hardness ~80 IRHD
Cable : Integral Cable, 3 m
Connector : 3 x 10-32 UNF
Mounting : Strapped, adhesive or pressed

Notes
The transverse sensitivity measurement is associated with approximately 4% uncertainty from electrical noise of the test equipment.
All values are typically at 25°C (77°F) unless measurement uncertainty is specified.


E&OE - specifications subject to change without notice.

Gracey & Associates   01933 624212   hire@gracey.com
British Standards accreditation for noise and vibration instrument hire and calibration
Home l Manufacturers l Applications l Hire Notes + FAQ l Site Search l Standards l Glossary

B&K Help Line - 01763 255780